ZENNIUM XC

Electrochemical Workstation : Highend Meets Compact


ZENNIUM XC
The ZENNIUM XC is a high-performance electrochemical workstation at a amazingly low-cost price. Besides a potentiostat/galvanostat it bears a frequency response analyzer unit ( FRA ) for promoting electric resistance spectroscopy. It provides high-end features like a frequency range up to 5 MHz, a soap. current of ±2.0 A, a controlled voltage of ±14 V and an input signal electric resistance of 10 TΩ || ±5 pF. The ZENNIUM XC is a arrant workstation for a wide range of electrochemical applications with stress on corrosion, coatings and laminates. The ZENNIUM XC can be easily switched from grounded to floating operation.

ZENNIUM XC comes bundled with the outstanding THALES XT software package which offers all standard methods and more at a mouse click. This is why the Zennium ninety can easily be adapted to very unlike measurement requirements. furthermore, with the probes available, the ZENNIUM XC is able to grow with its tasks :

  • Compliance Voltage up to ±120 V (CVB120)
  • Input Impedance up to 1000 TΩ (HiZ Probe)
  • FRA interface to 3rd-party electronic loads or potentiostats (FRA Probe)

Specifications

General
Overall bandwidth DC – 6 MHz
ADC resolution 32 bit
Harmonic
reject*
> 60 dB @ ½ full scale
Potentiostat modes potentiostatic, galvanostatic,
pseudo-galvanostatic, rest potential, ZRA, off
Cell connection 2-, 3-, 4-terminal kelvin
Ground reference grounded, floating
PC interface USB 2.0
Dimensions 255 x 160 x 385 mm
Weight 8 kg
Accessories 2 cell cable sets, USB cable, power cord, manual
Power supply 100/115/230 VAC, 50/60 Hz
Ambient temperature +10 °C to +30 °C
Ambient humidity < 60% without derating
Frequency generator & analyzer
Frequency range 10 µHz to 5 MHz
Accuracy* < 0.0025%
Resolution 0.0025%, 10000 steps/decade
Output potentiostatic
Controlled voltage   ±5 V
±14 V
Resolution 2.5 nV
7.5 nV
  5 V range
14 V range
Accuracy ±200 µV ± 10 ppm of reading
±600 µV ± 10 ppm of reading
  5 V range
14 V range
Integral Nonlinearity* typ. 4 ppm, max. 8 ppm
typ. 12 ppm, max. 24 ppm
  5 V range
14 V range
AC amplitude 0 to 2 V, 18 bit resolution
0 to 6 V, 18 bit resolution
  5 V range
14 V range
Bandwidth 6 MHz @ 33 Ω load
IR compensation auto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%
Small signal rise time 150 ns to 200 µs in 5 steps automatic
selection
Slew rate* 15 MV/s  
Phase shift* 10° @ 500 kHz  
Output
galvanostatic
Controlled current ±2 A
Current range ±1 nA to ±2 A in 12 steps Main
Resolution 32 bit, ±0.2 ppb of FS
(1ppm @ 1 pA)
Accuracy 0.04 % of FS ± 0.1 % of reading
0.2 % of FS ± 0.4 % of reading
≥ 1 µA to 100 mA
< 1 µA or > 100 mA
Input
Maximum
input
voltage
±5 V
±14 V
  5 V range
14 V range
Voltage
resolution
±2.5 nV
±7.5 nV
  5 V range
14 V range
Voltage
accuracy
  ±100 µV ± 5 ppm of reading
±300 µV ± 10 ppm of reading
  5 V range
14 V range
DC current
resolution
±5 aA (32 Bit)  Main / HiZ Probe
DC current
accuracy
± 0.08% of reading ± 0.04% of FS
± 0.8% of reading ± 0.4% of FS
± 0.8% of reading ± 0.4% of FS
± 0.5% of reading ± 125 fA
1 µA … 100 mA
100 mA … 2 A
10 nA … 1 µA
< |1 nA| (HiZ Probe)
Input
impedance
> 10 TΩ || ±5 pF (typical)
> 1000 TΩ || ±1 pF (typical)
Main
HiZ Probe
Input
leakage
current
< |± 200 fA| typ.
< 2 – 5 pA
< |± 10 fA| typ.
Main

HiZ Probe

Impedance
range
    1 mΩ to   10 GΩ /    2%
100 mΩ to   10 MΩ / 0.2%
100 mΩ to 100 GΩ /    2%
   10 µΩ to     1 GΩ /    2%
   1 mΩ to    10 MΩ / 0.2%
Main

HiZ
Gal

Common
mode
rejection
> 86 dB @   10 µHz to 100 kHz
> 66 dB @ 100 kHz to     5 MHz
 
Input
channel
phase –
tracking
accuracy
  ±0.05 ° @   10 µHz to 100 kHz
±0.125 ° @ 100 kHz to    5 MHz
Equivalent
effective
input
noise
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz Main

Zennium XC Flyer

Download the latest data aviator about ZAHNER Zennium XC potentiostat…
PDF-Document
Zennium ninety

Leave a Reply

Your email address will not be published.